61
If the
printer 26
is installed and enabled, but the printout is
not available at the end of the cycle, check that the
paper roll is positioned correctly and that the printer door
is properly closed. If the printer is out of paper or the door
is open, the ERROR light will flash on the printer.
•
Consult the PRINTER chapter and follow the
instructions provided.
If the POWER light on the
printer 26
is off, the printer is not
correctly powered:
•
contact a qualified technical service center.
The
auto water fill 29
(optional, standard on BHD models)
isn't working.
•
Check that the Self-filling option is set to ON in the
Main Settings Menu (see Ch. OPERATION).
•
If configuration A with external can is installed and
enabled, check whether there is water in the can.
If there is water and the SELF-FILLING ALARM is
active, check that the extraction tube is properly
positioned so that it touches the bottom of the can.
Check that the tube is not kinked or twisted. Check
that the height difference between the sterilizer
and the bottom of the can is not more than 1
meter (see CONNECTIONS, Ch. INSTALLATION).
•
If configuration B from the water mains is installed
and enabled, check that the water mains tap is
open. If the tap is open and the SELF-FILLING
ALARM is active, check that water is flowing
downstream from the Cominox demineralization
system.
•
If none of the solutions provided here solves the
problem, contact a technical service center, and
in the meantime disable the Self-Filling setting (see
Ch. OPERATION) and use MANUAL CLEAN WATER
REFILL (see Ch. INSTALLATION) to continue using the
sterilizer.
Steam loss from the
safety valve 12
.
•
If the sterilizer is installed at more than 1500 meters
above sea level, contact a technical service
centre to have a suitable safety valve installed (see
Ch. TECHNICAL SPECIFICATIONS).
•
If the altitude is below 1500 meters asl but the
problem persists, contact a technical service
center to check the efficiency of the valve.
AUTOMATIC WATER
FILL DOES NOT WORK
STEAM LOSS
FROM THE SAFETY
VALVE
Alarms
Содержание SterilClave 18 VLS
Страница 2: ...2...
Страница 7: ...7 Graphic representation of references Mod 18 Table of contents...
Страница 8: ...8 Graphic representation of references Mod 24 Table of contents...
Страница 9: ...9 06 06A OPTIONAL OPTIONAL Table of contents...
Страница 10: ...10...
Страница 70: ...70...
Страница 71: ...71...