65
Clean the
clean water tank 10
and the
waste water recovery tank 11
. Empty them
completely at least every six months and refill the clean water tank with clean water (see
Water supply table).
If the sterilizer will not be used for a long period, leave the tanks empty.
Filters
Replace the
air sterilization filter 15
at least
every one year
: remove the filter and screw in
the new one.
Annually
check the efficiency of the
waste filter 20
as follows:
- open the
door 16
.
- remove the filter by pulling upward.
- check it, and clean with a jet of compressed air or replace it.
- replace the filter.
Checks
At least
once a month
check the efficiency of the
door gasket 31
and replace it if there is
damage.
To extend the life of the filter, it is best to leave the
door 16
ajar, but not locked with the
handle, when the unit is not being used.
Every day
, check that there are no drips or seepage in the work area, since they can
cause a slipping danger.
Fuses
To calibrate the
fuses 32
, refer to the label plate inside the
door 18
.
At least annually, check the calibration of the pressure and temperature probes.
The probes check must be performed strictly by specialized personnel or at a
COMINOX TECHNICAL SERVICE CENTER
!
Maintenance
Содержание SterilClave 18 VLS
Страница 2: ...2...
Страница 7: ...7 Graphic representation of references Mod 18 Table of contents...
Страница 8: ...8 Graphic representation of references Mod 24 Table of contents...
Страница 9: ...9 06 06A OPTIONAL OPTIONAL Table of contents...
Страница 10: ...10...
Страница 70: ...70...
Страница 71: ...71...