57
ALARMS
The alarms indicate a need to conduct maintenance, or that some defect has been
found.
If a defect is encountered during a cycle, the cycle is interrupted immediately, and a
15-second beep indicates an alarm. The description of the alarm appears on the display,
including the cycle and phase in which it occurred. To reset the alarm, press SILENCE
ALARM and enter the security code.
Once the alarm has been reset and the cause behind it has been resolved if possible,
begin a new cycle. If the problem persists, contact an AUTHORIZED COMINOX TECHNICAL
SERVICE CENTER.
The load in the sterilizer during an interrupted cycle should be considered
NOT STERILE.
Description /
•
Solution
Activates in STAND BY (see CHECK WATER QUALITY, Ch.
INDICATORS). This indicates that the water conductivity
value has exceeded the acceptable level.
•
If a Cominox (Speedy Water or Osmosis) water
demineralization system is installed, change the
cartridge.
•
If the water is filled from an external can, check
that water.
Indicates that distilled or demineralized water is lacking in
the
clean water tank 10.
•
If the alarm is triggered at the beginning of a cycle,
fill the tank (see MANUALLY FILL CLEAN WATER
TANK, Ch. INSTALLATION).
•
If the alarm is triggered during a cycle, fill the tank
or follow the instructions provided for the
INSUFFICIENT STEAM ALARM.
Indicates that the necessary time was exceeded for the
correct water level to be reset in the
chamber 09
.
•
Recheck the tilt of the sterilizer (see Ch.
INSTALLATION).
Indicates that there is a leak of steam or overheated
steam in the
chamber 09
.
•
Check the efficiency and cleanliness of the
door
gasket 31
as well as the internal
door 16
disk the
seal is fitted to. Replace if damaged (see
CLEANING and CONTROLS in Ch. MAINTENANCE).
•
Run a VACUUM test to check for pressure leaks.
•
Recheck the tilt of the sterilizer (see Ch.
INSTALLATION).
•
Check that the total maximum load weight (see
Table of cycles/sterilizable material) has not been
exceeded.
INSUFF. WATER
QUALITY ALARM
INSUFF. WATER LEVEL
ALARM
CHAMBER LEVEL
ALARM
INSUFFICIENT
STEAM ALARM
Alarms
Содержание SterilClave 18 VLS
Страница 2: ...2...
Страница 7: ...7 Graphic representation of references Mod 18 Table of contents...
Страница 8: ...8 Graphic representation of references Mod 24 Table of contents...
Страница 9: ...9 06 06A OPTIONAL OPTIONAL Table of contents...
Страница 10: ...10...
Страница 70: ...70...
Страница 71: ...71...