50

EVENTS in the CYCLE 

Press EVENT to access the cycle events screen: 

 

     

 

 

      

 

 

The display will show information about the phases of the cycle in progress: the phase 
name, the time the phase began, the temperature, and the pressure of the sterilization 
chamber at the beginning of the phase. For the sterilization phase, the display also shows 
the minimum and maximum values of the temperature, pressure, and theoretical steam 
temperature. 

Use the 

↑↓

 arrows to scroll through the full list, or press the 

Å

 button to exit. Press STOP to 

interrupt the cycle. 
The upper part of the display shows the icons corresponding to the Door Locked indicator 

and the Cycle Completed indicator when the cycle ends. 

 
 

Operation 

Содержание SterilClave 18 VLS

Страница 1: ...for strictly complying with the instructions and information it provides COMINOX is not liable for any damage to people things or the sterilizer itself in the event that the operator fails to comply w...

Страница 2: ...2...

Страница 3: ...cles sterilizable material 24B 24BHD 15 Table of cycles sterilizable material 24S 15 TECHNICAL SPECIFICATIONS 17 Technical specifications table 18 17 Technical specifications table 24 18 REFERENCE STA...

Страница 4: ...2 5 Self filling 33 2 6 Printer 33 2 7 Cycle Recording 33 2 8 Data Logging 33 3 Advanced Tools Menu 34 3 1 Prog SPECIAL 34 3 2 Programmed start 35 3 3 Hollow Loads 36 3 4 Code option 37 3 4 1 Create U...

Страница 5: ...62 MAINTENANCE 64 GENERAL CONDITIONS 64 Cleaning 64 Filters 65 Checks 65 Fuses 65 STORAGE AND DISPOSAL 66 Sterilizer storage or long term disuse 66 Disposal and scrapping 66 Maintenance schedule table...

Страница 6: ...ter tank 11 waste water recovery tank 12 safety valve 13 main switch 14 filler 15 air sterilization filter 16 door 17 safety thermostat 18 print door 19 coil covering grid 20 waste filter 21 chamber p...

Страница 7: ...7 Graphic representation of references Mod 18 Table of contents...

Страница 8: ...8 Graphic representation of references Mod 24 Table of contents...

Страница 9: ...9 06 06A OPTIONAL OPTIONAL Table of contents...

Страница 10: ...10...

Страница 11: ...roper use of the sterilizer changes that occur during transportation or those caused by specific environmental conditions lack of maintenance or maintenance conducted improperly or problems due to tam...

Страница 12: ...ce Manual each sterilizer is supplied with a Guarantee Certificate and the EC Declaration of Conformity The Cominox SterilClave sterilizer also includes the following accessories Supplied as standard...

Страница 13: ...he operator in selecting the sterilization cycle see Table of cycles sterilizable material Strictly comply with the sterilization instructions and information specified by the manufacturer for any mat...

Страница 14: ...no yes yes yes Maximum load weight solid hollow porous 5 Kg 5 Kg 3 Kg 2 5 1 5 Kg 2 5 1 5 Kg 2 5 1 5 Kg Table of cycles sterilizable material 18S 121 UN WRAPPED 134 UN WRAPPED 134 WRAPPED 121 POROUS HO...

Страница 15: ...Maximum load weight solid hollow porous 7 5 Kg 7 5 Kg 4 Kg 3 7 2 2 Kg 3 7 2 2 Kg 3 7 2 2 Kg Table of cycles sterilizable material 24S 121 UN WRAPPED 134 UN WRAPPED 134 WRAPPED 121 POROUS HOLLOW 134 P...

Страница 16: ...d where the length of the cavity is no longer than 3000 mm L 3000 mm TYPE B HOLLOW LOADS instruments with an open cavity on one side only where the length to cavity diameter ratio is greater than or e...

Страница 17: ...pressure maximum kPa 331 Electrical voltage and frequency V Hz 230 50 60 single phase AC Maximum allowed voltage fluctuations 10 Absorbed maximum power W 2000 Heat emission J h 0 3 Detected noise leve...

Страница 18: ...l tank and maximum load kg cm 1 51 1 53 1 63 Room temperature allowed C from 15 to 40 Maximum relative humidity allowed 85 Maximum altitude allowed with standard safety valve m asl 1500 2000 with repl...

Страница 19: ...t 395 475 Depth 560 620 Usable space dimensions SterilClave 18 SterilClave 24 Width 185 185 Height 150 150 Depth 285 420 NEVER place the sterilizer on soft surfaces such as cloth or foam and never rem...

Страница 20: ...e any sterilizer components It is strictly forbidden to clean or wash the sterilizer with jets of water Never place containers of liquids on the sterilizer or the shelves above it Leakage or spillage...

Страница 21: ...organisms Dispose of according to the regulations in force Escaping steam and or gas fumes when opening the door 16 Always carefully check whether the load complies with cycle temperatures and open th...

Страница 22: ...same applies to any device used to remove packaging scissors hammers tongs knives etc Packaging components must be removed and disposed of according to the regulations in force in each country You ar...

Страница 23: ...tion site it must be stored in a dry protected location If you wish to cover the sterilizer with plastic sheeting you must first insert silica gel or other desiccant systems inside the packaging Ideal...

Страница 24: ...are kept under or above any live equipment in unsuitable built in units without the fitted adjustable feet in domestic environments accessible to untrained personnel on surfaces that are flammable sof...

Страница 25: ...the power supply simply plug in the sterilizer by inserting the power cable 03 plug into a suitable socket see GENERAL CONDITIONS and the Technical specifications table Water connection This option c...

Страница 26: ...ry tank 11 will start to drain Water supply table Water supply Residue from evaporation 10 mg l Silicon Dioxide SiO2 1 mg l Iron 0 2 mg l Cadmium 0 005 mg l Lead 0 05 mg l Remaining heavy metals exclu...

Страница 27: ...te pipe 02 provided to the waste water tap 25 after placing a container of sufficient size under the other end see Technical specifications table to collect the water Open the tap to allow any water l...

Страница 28: ...he display the display will show another screen that asks for the personal security code Pressing the key deletes the last number entered or if no numbers are shown on the display it returns to OFF mo...

Страница 29: ...ates that the conductivity value of the water introduced has exceed the optimum level but is still acceptable the sterilizer allows a cycle to begin but it is advisable to follow the instructions prov...

Страница 30: ...pipe is not bent or clogged MENU From STAND BY using the four multi function buttons you can navigate the menu and program the various SterilClave functions When you select MENU from STAND BY the opti...

Страница 31: ...HOLLOW POROUS cycle at 134 C for solid porous and hollow products packaged or unpackaged 6 134 PRION cycle at 134 C for solid porous and hollow products that have come in contact with BSE infective a...

Страница 32: ...the arrows to change the value Press ENTER to confirm the information and move on to the next setting Press the key to exit without changing the setting value 2 3 Button Sound Press ENTER to show the...

Страница 33: ...ER to confirm the information or press to exit without changing the value of the setting 2 7 Cycle Recording This item is only available if the sterilizer is equipped with a USB Port 06 optional stand...

Страница 34: ...f the SPECIAL cycle Press ENTER to change the values of the SPECIAL cycle parameters In particular the sterilization temperature from 117 C to 134 C the sterilization duration The maximum value is 30...

Страница 35: ...om among 121 UNWRAPPED 134 UNWRAPPED 134 WRAPPED 121 HOLL POROUS 134 HOLL POROUS 134 PRION SPECIAL USER BOWIE DICK TEST VACUUM TEST VACUUM B D TEST The VACUUM B D TEST cycle calls for the Vacuum Test...

Страница 36: ...rammed start and return to STAND BY 3 3 Hollow Loads This function allows for automatic recognition of the load type HOLLOW A HOLLOW B or SOLID entering the length and diameter of the cavity in millim...

Страница 37: ...ress ENTER the type of load will appear on the display Press the button to exit 3 4 Code option This option is only available if the user connected is the Administrator There are two code management f...

Страница 38: ...ss the button to delete the last number entered or if there are no numbers on the display to exit Press ENTER to access the screen for entering the name of the new user Using the keys scroll through t...

Страница 39: ...he screen for setting the values for the SPECIAL cycle parameters If user programming is ON the new user will be allowed to change the parameters of the SPECIAL cycle If it is set to OFF the new user...

Страница 40: ...te User Press ENTER to display the list of users with their access code Scroll the list with the arrows to select the user to be deleted and press ENTER Press YES to delete the user code press NO to e...

Страница 41: ...R to display 4 1 BACK UP Select this option to copy the data from the internal memory of the sterilizer to the USB key The user can chose the data to transfer the cycles data of the last month all the...

Страница 42: ...e 4 multi function buttons will return the sterilizer to STAND BY mode 4 2 Cycle records Press ENTER to display The user can chose to display the cycles data per day or all the cycles with an alarm Af...

Страница 43: ...arm a description of the alarm will be shown with the values for cycle duration temperature and pressure when the alarm was triggered PRINT is only enabled if the sterilizer features a printer 26 opti...

Страница 44: ...led maintenance For each component the remaining days and cycles are shown along with the last three replacements the date of the operations and the total number of cycles at that point Use the arrows...

Страница 45: ...Only authorized technicians can replace and record the other items 5 2 Save Maintenance Press ENTER to access this screen Press NO to return to the Maintenance Menu Press YES to save the maintenance...

Страница 46: ...lizer Use the arrows to scroll through the information and view Relative pressure Theoretical steam temperature Total number of cycles Water conductivity of last load only if automatic fill is availab...

Страница 47: ...he cycle in progress The temperature in the sterilization chamber The pressure in the sterilization chamber The cycle duration The end phase time for sterilization and drying phases The Door locked in...

Страница 48: ...ilization cycle When the STOP button is pressed the sterilizer shows a MANUAL STOP ALARM describing the cycle and the phase in which the alarm was triggered The alarm can be silenced by pressing any o...

Страница 49: ...has not been damaged e g open seals Material in damaged packages must be repackaged and run through a new sterilization cycle If installed and enabled the printer 26 optional standard on PRINT models...

Страница 50: ...amber at the beginning of the phase For the sterilization phase the display also shows the minimum and maximum values of the temperature pressure and theoretical steam temperature Use the arrows to sc...

Страница 51: ...rogramming file Use the arrows to scroll through the full list or press to exit Press STOP to interrupt the cycle The upper part of the display shows the icons corresponding to the Door Locked indicat...

Страница 52: ...e lower tray close the door 16 press ON and enter the access code press the MENU key from STAND BY select the Menu Cycle number 1 and press ENTER use the arrows to select the Bowie Dick test cycle Pre...

Страница 53: ...the display the test will be interrupted automatically Before repeating the test open the door 16 let the chamber 09 cool completely and check that it is perfectly dry Whenever you want to run the VAC...

Страница 54: ...e and open the door 16 remove the tray holding the test and check the results of the test according to the manufacturer s instructions Printer If installed and enabled see Main Settings Menu sec 2 6 a...

Страница 55: ...Port 06 optional standard on USB models and the Cycle Recording option has been enabled in the Main Settings Menu see sec 2 7 at the end of the cycle the information will be saved automatically in the...

Страница 56: ...is still acceptable the sterilizer allows a cycle to begin but it is advisable to proceed according to instructions given for the WATER QUALITY INSUFFICIENT ALARM see ALARMS Indicates that the waste...

Страница 57: ...If the water is filled from an external can check that water Indicates that distilled or demineralized water is lacking in the clean water tank 10 If the alarm is triggered at the beginning of a cycle...

Страница 58: ...16 disk the seal is fitted to Replace if damaged see Ch MAINTENANCE Check that the drain filter 20 is clean and positioned correctly see Ch MAINTENANCE Run a VACUUM test to check for pressure leaks Ch...

Страница 59: ...or Indicates that during draining pressure did not drop to the correct preset value Follow the instructions provided for the INSUFFICIENT VACUUM ALARM Check that the total maximum load weight see Tabl...

Страница 60: ...is still water in the chamber 09 Check that the drain filter 20 is clean and positioned correctly see Ch MAINTENANCE Recheck the tilt of the sterilizer see Ch INSTALLATION Drying after the cycle is n...

Страница 61: ...k that the tube is not kinked or twisted Check that the height difference between the sterilizer and the bottom of the can is not more than 1 meter see CONNECTIONS Ch INSTALLATION If configuration B f...

Страница 62: ...RM 08 CHAMBER LEVEL ALARM 09 MANUAL STOP ALARM 10 INSUFFICIENT VACUUM ALARM 11 FRAC VACUUM ALARM 12 13 PROBE LEVEL ALARM 14 PRESSURIZATION ALARM 15 OVERTEMP BAND ALARM 16 UNDERTEMP BAND ALARM 17 18 PO...

Страница 63: ...en switch to OFF mode Pressing ON with one of the four multi function buttons will show Press NO to enter the STAND BY mode with door locked Press YES to start the door unlocking procedure When the do...

Страница 64: ...found use stainless steel polishing products Then remove any product residue with a degreasing product and rinse thoroughly with distilled water Clean the workplace every day checking that there are n...

Страница 65: ...ck it and clean with a jet of compressed air or replace it replace the filter Checks At least once a month check the efficiency of the door gasket 31 and replace it if there is damage To extend the li...

Страница 66: ...COMINOX or an AUTHORIZED TECHNICAL SERVICE CENTER for more information When the unit will not be used for a long time empty the tanks completely Disposal and scrapping N B COMINOX declines all respons...

Страница 67: ...ars Vacuum pump kit 1500 2 years EV2 drain 1500 2 years EV6 by pass 1500 2 years Drain tank probe 1500 2 years Water fill pump 3000 3 years Internal coil 6000 6 years Door gasket 1500 2 years Chamber...

Страница 68: ...68 APPENDIX 1 GRAPHICS 18 B 18 BHD 24B 24 BHD 121 UNWRAPPED 134 UNWRAPPED 121 HOLLOW POROUS 134 HOLLOW POROUS 134 PRION 134 WRAPPED Appendix 1...

Страница 69: ...69 APPENDIX 2 GRAPHICS 18 S 24 S 121 UNWRAPPED 134 UNWRAPPED 121 HOLLOW POROUS 134 HOLLOW POROUS 134 PRION 134 WRAPPED Appendix 2...

Страница 70: ...70...

Страница 71: ...71...

Страница 72: ...20841 Carate B za MB Italia tel 39 0362 912312 Fax 39 0362 900940 www cominox it UCG02VLS May 2014 We reserve the right to modify the products without any notice In the event of disputes the relevant...

Отзывы: