58
Indicates that small steam leaks or an unsaturation of
steam has prevented the correct rise in temperature in the
chamber 09
during the pressurization phase.
•
Follow the instructions specified in the case of the
INSUFFICIENT STEAM ALARM.
•
If the alarm persists, this could be due to a problem
in the electric heating circuit: contact a technical
service center.
Indicates that the temperature inside the
chamber 09
has
exceed 150°C.
•
If the alarm persists, contact a technical service
center.
Indicates that, during a vacuum peak in the vacuum
pulse phase, the pressure failed to drop to the preset
value.
•
Follow the instructions specified in the case of the
INSUFFICIENT VACUUM ALARM.
•
Check that the total maximum load weight (see
Table of cycles/sterilizable material) has not been
exceeded.
Indicates that during the pre-vacuum, the pressure failed
to drop to the preset value.
•
Check the efficiency and cleanliness of the
door
gasket 31
as well as the internal
door 16
disk the
seal is fitted to. Replace if damaged (see Ch.
MAINTENANCE).
•
Check that the
drain
filter 20
is clean and
positioned correctly (see Ch. MAINTENANCE).
•
Run a VACUUM test to check for pressure leaks.
•
Check that ventilation air vents are clear (see PREP.
FOR BUILT-IN, Ch. INSTALLATION) and ask a
technician to check the efficiency of the internal
fan(s).
STERILIZATION TEMPERATURE BAND: OVERTEMP.
Indicates that during sterilization the temperature or
theoretical steam temperature
has exceeded the
maximum temperature anticipated for the sterilization
temperature range.
•
If the alarm persists, contact a technical service
center to check the calibration of the temperature
and pressure sensors.
STERILIZATION TEMPERATURE BAND: UNDERTEMP.
Indicates that during sterilization, the temperature or
theoretical temperature of the steam
has dropped below
the minimum value anticipated for the sterilization
temperature range.
•
Check that the total maximum load weight (see
Table of cycles/sterilizable material) has not been
exceeded.
PRESSURIZATION
ALARM
PT1 OVERTEMP.
ALARM
FRAC. VACUUM
ALARM
INSUFFICIENT VACUUM
ALARM
STERIL. TEMP. BAND:
OVERTEMP.
STERIL. TEMP. BAND:
UNDERTEMP.
Alarms
Содержание SterilClave 18 VLS
Страница 2: ...2...
Страница 7: ...7 Graphic representation of references Mod 18 Table of contents...
Страница 8: ...8 Graphic representation of references Mod 24 Table of contents...
Страница 9: ...9 06 06A OPTIONAL OPTIONAL Table of contents...
Страница 10: ...10...
Страница 70: ...70...
Страница 71: ...71...