59
Indicates a problem with the level probes.
•
Empty and refill the
clean water tank 10
when the
unit is off and idle.
Indicates that the
chamber 09
is overheated, and the
safety thermostat 17
has been triggered.
•
Contact a technical service center.
Activates in STAND BY when the
auto water
fill 29
is
installed (see Ch. CONNECTIONS).
•
If configuration A is set, the distilled or
demineralized water tank must be refilled or
replaced.
•
If configuration B is set, check that the water main
tap is open and that the mains pressure is
compliant.
Indicates that the
door 16
was not closed correctly.
•
Close the door firmly before starting a cycle.
This alarm triggers when the temperature sensor in the
sterilizer is disconnected or broken.
•
Contact a technical service center to check the
temperature sensor.
Indicates that during draining, pressure did not drop to the
correct preset value.
•
Follow the instructions provided for the INSUFFICIENT
VACUUM ALARM.
•
Check that the total maximum load weight (see
Table of cycles/sterilizable material) has not been
exceeded.
Indicates that the temperature in the
chamber 09
does
not correspond to the theoretical steam temperature.
•
If the alarm persists, contact a technical service
center to check the calibration of the temperature
and pressure sensors.
This alarm triggers when the pressure sensor in the sterilizer
is disconnected or broken.
•
Contact a technical service center to check the
pressure sensor.
Indicates the while the cycle was underway, the STOP
button was pressed before the natural completion of the
cycle.
•
Consider the load to be NOT STERILE and run a new
sterilization cycle.
DRAIN
ALARM
MISALIGNMENT
ALARM
PRESSURE TRANSD.
ALARM
COILS
ALARM
SELF-FILLING
ALARM
DOOR
ALARM
MANUAL STOP
ALARM
TEMP. SENSOR PT1
ALARM
LEVEL PROBES
ALARM
Alarms
Содержание SterilClave 18 VLS
Страница 2: ...2...
Страница 7: ...7 Graphic representation of references Mod 18 Table of contents...
Страница 8: ...8 Graphic representation of references Mod 24 Table of contents...
Страница 9: ...9 06 06A OPTIONAL OPTIONAL Table of contents...
Страница 10: ...10...
Страница 70: ...70...
Страница 71: ...71...