48
STOP the CYCLE
When the cycle is complete, press STOP or open the
door 16.
The sterilizer asks for the personal access code and then returns to STAND BY mode.
The STOP button will halt the cycle in progress.
If a cycle is interrupted with a load of packaged instruments, even when the sterilization
phase is complete, sterility is not guaranteed because the load has not dried. Unless they
are reused immediately, the instruments should be run through another sterilization cycle.
When the STOP button is pressed, the sterilizer shows a MANUAL STOP ALARM, describing
the cycle and the phase in which the alarm was triggered.
The alarm can be silenced by pressing any one of the four multi-function buttons:
The sterilizer will ask to enter the security code and then switch to OFF mode.
Pressing ON with one of the four multi-function buttons will show:
You can interrupt the cycle at any time by pressing STOP, but you are advised to
stop the cycle only in an emergency.
!
When the STOP button is pressed, the door locking device will still be engaged.
The display will show the icon “Door locked” and the door cannot be opened.
!
Operation