27
MANUALLY FILL CLEAN WATER TANK
If Self-filling option is not available or not selected (see WATER CONNECTION), fill the tank
manually.
Unscrew and remove the cap from the
filler 14
, insert the
metering funnel 05
provided,
and pour in distilled or suitably demineralised water (see Water supply table and Technical
specifications table for the quantity).
If you fill the sterilizer when it is switched on and in STAND BY mode, the MIN WATER LEVEL
and MAX WATER LEVEL indications on the display will allow you to check the level. An
audible signal lasting three seconds will go off when the MAX WATER LEVEL indication
appears, indicating that the tank is full.
Top up with water whenever the MIN WATER LEVEL indicator appears on the display.
For maximum operating time before refilling, fill the tank completely until the MAX WATER
LEVEL indicator appears, and completely empty the
waste water recovery tank 11
(see
MANUALLY DRAIN TANKS).
MANUALLY DRAIN TANKS
When there is no DIRECT DRAIN (see WATER CONNECTION) or it is not connected, the
waste water recovery tank 11
must be drained manually whenever the DRAIN WASTE
WATER indicator appears on the display (see SIGNALS AND ALARMS).
Connect the
waste pipe 02
provided to the
waste water tap 25
after placing a container
of sufficient size under the other end (see Technical specifications table) to collect the
water.
Open the tap to allow any water left in the
waste water recovery tank 11
to drain.
For maximum operating time before refilling, empty the tank completely and completely
fill the
clean water tank 10
until the MAX WATER LEVER indicator appears (see MANUALLY
FILL CLEAN WATER TANK).
The
clean water drainage tap 24
is used to empty the
clean water tank 10
for
maintenance or transport (see Chap. TRANSPORTATION and MAINTENANCE) and it is only
used under these conditions.
PREPARING FOR BUILT-IN INSTALLATION
If the sterilizer will be built into a furniture item or another storage rack, you must follow ALL
the instructions provided for normal installation, plus you must:
- comply with the built-in distances (see Chap. DIMENSIONS).
- provide adequate ventilation for the components.
- take into account
heat emissions
produced by the sterilizer (see Technical specifications
table).
- allow for the
door 16
to open.
- allow for routine maintenance.
- prevent damage to the sterilizer.
-
provide for the room necessary for installing the accessories for a water supply fitting.
WARNING: the temperature of the liquid can reach 80°C and the liquid may be
contaminated. The liquid should be disposed of according to the legislation in
force
.
!
Installation
Содержание SterilClave 18 VLS
Страница 2: ...2...
Страница 7: ...7 Graphic representation of references Mod 18 Table of contents...
Страница 8: ...8 Graphic representation of references Mod 24 Table of contents...
Страница 9: ...9 06 06A OPTIONAL OPTIONAL Table of contents...
Страница 10: ...10...
Страница 70: ...70...
Страница 71: ...71...