INSTALLATION
GENERAL CONDITIONS
Installation of the sterilizer is to be carried out according to Customer requirements and
those of the site where the sterilizer is to be installed. Therefore, it is important that the
Customer promptly informs
COMINOX
of any special situations (such as ambient
conditions, size, or built-in limits).
In general, it is necessary to:
- check that any ambient conditions are acceptable (explosion risk, excessive ventilation,
or high humidity).
- check that the unit is not exposed to any adverse weather conditions, such as sun, rain,
wind, etc.
- check that the power supply complies with the specifications provided on the label
plates affixed to the sterilizer.
- check that the power supply is properly grounded.
Arrangement
The sterilizer should be installed in a laboratory or medical or dental office where it can be
accessed exclusively by
authorized personnel
.
IT IS NOT BE INSTALLED:
- near sources of heat.
- in rooms where volatile explosive substances are kept.
- under or above any live equipment.
- in unsuitable built-in units.
- without the fitted adjustable feet.
- in domestic environments accessible to untrained personnel.
- on surfaces that are flammable, soft, unstable, or have wheels.
Position the unit in the area where it will be installed, on a sturdy, perfectly horizontal
surface.
When installing the unit for the first time (and whenever it is removed from its original
installation location), follow the procedure indicated below:
- position the sterilizer where it will be installed and do not move it thereafter.
- open the
door 16
and empty the
chamber 09
.
- remove the
coil covering grid 19,
loosening the corresponding screw.
- remove the
tray support 08
.
- remove the
waste filter 20
and place the supplied
level testing pipe 27
in the hole. The
filter is located in the middle at the bottom of the chamber.
Installation
The operation must be performed by trained, qualified personnel only: it is
important to comply with the specifications in this manual and follow the
technical diagrams and documents.
!
Avoid any electro-magnetic interference that could affect the proper operation
of electronic equipment, leading to consequent dangers.
!
Содержание SterilClave 18 VLS
Страница 2: ...2...
Страница 7: ...7 Graphic representation of references Mod 18 Table of contents...
Страница 8: ...8 Graphic representation of references Mod 24 Table of contents...
Страница 9: ...9 06 06A OPTIONAL OPTIONAL Table of contents...
Страница 10: ...10...
Страница 70: ...70...
Страница 71: ...71...