Technical alarms and status messages
CNAP
®
Monitor 500
Page 14-5
Version 5.2.5,
h
21-FHZU-10002
Message
Possible cause
Measures
CNAP: Calibrating NBP
-
NBP measurement to cali-
brate CNAP
®
blood pressure
is in progress
-
Wait for end of NBP mea-
surement
CNAP: Artifact
-
CNAP
®
blood pressure is not
within physiological measur-
ing range
-
Low signal amplitude in
CNAP
®
double finger cuff
-
Interference because of third
party measuring devices
-
CNAP
®
hardware is ringing
due to artifacts
-
Check and eliminate influ-
ence from third party mea-
suring devices
-
Avoid artifacts (e.g. move-
ments)
-
Check CNAP
®
cables and
connectors
-
Check CNAP
®
double finger
cuff
-
Replace CNAP
®
double finger
cuff and cable
CNAP:
Cuff Expiring
-
CNAP
®
double finger cuff is
reaching end of lifecycle,
thus providing low quality of
measurement (remaining
lifetime less than 12 hours)
-
Replace CNAP
®
double finger
cuff
CNAP: Cuff Ambient
Light
-
Ambient light interferes with
CNAP
®
double finger cuff
-
Reduce ambient light (i.e.
brightness, switch off, …)
-
Check setup of CNAP
®
double
finger cuff
CNAP: CO Calibration
Failed
-
Manual CO calibration not
possible due to signal arti-
facts
-
Reduce artifacts by perform-
ing a manual change finger
or repositioning of the CNAP
®
double finger cuff
CNAP: CNAP cuff is too
big
-
A too big finger cuff was
used on the patient
-
Use a smaller finger cuff on
the patient
-
Check the correct placement
of the finger cuff
CNAP: CNAP cuff is too
small
-
A too small finger cuff was
used
-
Use a bigger finger cuff on
the patient
-
Check the correct placement
of the finger cuff
14.4.2
Technical alarms
Message
Priority
Possible cause
Measures