Alarm limits – factory settings
The CNAP
®
Monitor 500 has been preset to the following factory settings and default settings for
alarm limits, which apply to both CNAP
®
and NBP.
•
Alarm limits (ADULT):
Lower limits
Upper limits
Parameter
Lower limits
Defaults
Upper limits Lower limits
Defaults
Upper limits
sBP
[mmHg]
40
90
255
45
140
260
dBP
[mmHg]
30
50
245
35
90
250
mBP
[mmHg]
35
60
250
40
110
255
Pulse
[bpm]
30
50
195
35
110
200
•
Alarm limits (PEDIATRIC):
Lower limits
Upper limits
Parameter
Lower limits
Defaults
Upper limits Lower limits
Defaults
Upper limits
sBP
[mmHg]
40
70
175
45
120
180
dBP
[mmHg]
30
40
165
35
70
170
mBP
[mmHg]
35
50
170
40
90
175
Pulse
[bpm]
30
75
195
35
130
200