Management of patient data
CNAP
®
Monitor 500
Page 5-3
Version 5.2.5,
h
21-FHZU-10002
5.2
Editing of patient data
At any given time – except during the display of the
Setup Patient
dialog – you can enter de-
tailed patient data by using the click-wheel control to select the
Alarm frame
on the main screen
and to open the
Patient Data
menu.
If the option
“Hemodynamic”
is enabled in the menu (see chapter 12) – with device type
CNAP
®
Monitor 500i/at+HD – the menu
Patient Data
is displayed automatically after starting a
measurement to enter the data. The
Patient Data
menu can also be manually opened and edited
via the “Hemodynamic parameter frame” (see chapter 3.4.5.) at any time (also during ongoing
measurements).
Menu item
Description
Settings
Category
Patient category:
ADULT > 14 years
PEDIATRIC 4 – 14 years
Pediatric, Adult
Name
Patient's surname and first name
(max. 20 characters)
Keyboard (click-
wheel control)
ID#
Patient file number, e.g. 12345678
(max. 15 characters)
Keyboard (click-
wheel control)
Gender*
Patient gender
---, M, F
Age*
Patients age in yeas
4 – 99
Weight*
Weight of patient in kg
20 – 200
Height*
Patient’s body height in cm
100 - 220
Discharge
Discharge patient information
Yes, No
OK
*
Required information for the calculation of the hemodynamic parameters – see chapter 12
STOP:
•
Patient category
:
Entering the correct patient category is an indispensable prere-
quisite before starting a measurement process. Be sure to select the correct patient
category as this determines the adjustment of alarm limits and the inflation pressure
of the NBP cuff.
NOTE:
•
Before applying the CNAP
®
finger cuff and during a measurement, the
Patient Da-
ta
menu
can be selected via the
Alarm frame
.
•
If, before starting a measuring process, you want to enter additional patient data,
you have to do this before applying the CNAP
®
finger cuff. By selecting the
Use
Current Patient Data
option in the
Setup Patient
dialog, the currently entered
patient data will be used.