Monitor configuration
CNAP
®
Monitor 500
Page 4-11
Version 5.2.5,
h
21-FHZU-10002
Procedure for activating data recording on the USB device:
NOTE:
•
There are two options to safely remove the USB stick after data recording:
1.
Power off CNAP
®
Monitor 500 first and then remove USB stick.
2.
After end of measurement set menu entry
Record
to
Off
in the menu
Se-
tup | Monitor | Record
. The USB stick is released and can be safely re-
moved.
•
If a USB stick has been removed improperly, it can only be re-integrated in the
CNAP
®
Monitor and re-used for further data recording by re-starting the device.
1. Prepare a USB stick with the following specifications:
•
‘Corsair Flash Voyager’ 4GB recommended
•
Writable
•
FAT32 formatted (not NTFS)
•
Old data deleted
2. Turn off the CNAP
®
Monitor 500
3. Connect USB stick
4. Turn on the CNAP
®
Monitor 500
5. Open the monitor menu in
Setup
/Monitor
and select one of the following settings for record-
ing:
Setting
Use
Parameters
Capacity
CSV File
Clinical focus: Data storage in
the CSV format (comma se-
parated values)
CNAP
®
(wave form and
beat-to-beat values),
NBP, pulse rate, time,
interventions, measure-
ment parameters
approx. 10 MB/h
(4GB = 400h)
Advanced
Additional storage of mea-
surement data in PDP file
format (for internal use only)
Further technical para-
meters
approx. 180 MB/h
(4GB = 22h)
6. Back to
Main Screen
7. Check if the USB stick properties are displayed in the top right of the screen.
USB Rec. - USB device connected
Used: x % - Used storage capacity in %U:
Stored capacity
In the event of failed activation, status messages will be displayed (see 14.1.1 – Status Messages).
Distributed by BIOPAC Systems, Inc.
Manufactured by CNSystems AG