Alarm system
CNAP
®
Monitor 500
Page 6-3
Version 5.2.5,
h
21-FHZU-10002
Illustration 28: Alarm frame – alarm conditions
Parameter frame
:
•
Background color: YELLOW – Physiological alarms (medium priority)
CYAN - Technical malfunction alarms (low priority)
BLUE - Status messages
•
Flashing parameters which are exceeding the alarm limits
•
Numeric values: unchanged during physiological alarms, blanked during technical malfunction
alarms by 3 stars (
***
).
Illustration 29: Parameter frame – alarm conditions
6.2
Acoustic alarm signals
In accordance with the regulations of IEC 60601-1-8, the CNAP
®
Monitor 500 produces acoustic
alarm signals. The differently coded alarm signals are repeated until acknowledged by pressing the
Alarm Pause/Off
key
.
NOTE:
Repetition rate for acoustic alarm signals is:
•
5 seconds for physiological alarms,
•
18 seconds for technical malfunction alarms.
STOP:
•
Do not solely rely on the acoustic alarm signals! Especially if the alarm volume is set
low or has been turned off, alarms might be missed, which could constitute a possible
danger for patients!
The alarm signal volume is individually adjustable. Factory setting is 80% of maximum volume and
can be adjusted from 20% to 100%.
The sound pressure at one meter distance to the device is defined for the different acoustic alarm
signals within the following ranges:
** CNAP: Sys High
** CNAP: Sys High
Distributed by BIOPAC Systems, Inc.
Manufactured by CNSystems AG