Alarm system
CNAP
®
Monitor 500
Page 6-4
Version 5.2.5,
h
21-FHZU-10002
Alarm Type
Volume 20%
1
Volume 100%
1
Tone Sequence
2
Physiological
54 dB
83 dB
B4 – D#5/Eb5 – F#5/Gb5
Technical
54 dB
79 dB
D#5/Eb5 – B4
Information Signal
34 dB
67 dB
C5
Reminder Signal
54 dB
78 dB
B4
1
A-weighted sound pressure level averaged over the measurement surface (dB)
2
Scientific notation
6.3
Alarm system control
The fast access key
Alarm Pause/Off
has various functions. The following table summa-
rizes all possible functions:
Menu item
Press
button
Alarm -
Reminder
Alarm-
Pause
Function
Acknowledgement
indefinite
1x
Off
N/A
All active audio alarms off
Acknowledgement
timed
1x
1 min, 2 min
or 3 min
N/A
All active audio alarms off for
the duration of the selected
period
Alarm Off
2x
N/A
No timeout
All physiological alarms are
deactivated
Alarm Pause
2x
N/A
1 min, 2
min or 3
min
All physiological alarms are
paused for the duration of the
selected period
Alarm Active
3x
N/A
N/A
All alarms are re-activated
after the Alarm Pause/Off
6.3.1
Acknowledgement of alarms
In order to acknowledge all activated alarms (physiological and technical malfunction alarms),
press
Alarm Pause/Off
once.
Depending on the respective settings of the
Alarm Reminder
feature, an acoustic reminder
signal is activated according to a predetermined time period or the alarm reminder remains dis-
abled until the end of this alarm condition.
Alarm Reminder:
If the alarm reminder is activated in the monitor setup, a repeated acoustic
signal reminds the operator of the alarm conditions that continue to exist after the acknowledge-
ment of the alarm signal by the operator. This acoustic reminder may be repeated for a limited or
unlimited amount of time.
Menu item
Description
Settings
Alarm Reminder
Setting of alarm reminder
off, 1 min, 2 min, 3 min
NOTE:
•
During measurements, an alarm reminder setting may be entered in the
Parame-
ter
menu by using the click-wheel control to open the
Parameter frame
. The set-
tings are saved by confirming with
Alarm Pause/Off
.
•
The alarm reminder operator setting may be set in the
Alarm defaults
menu,
which is opened from the
Service
menu by using the click-wheel control.