Alarm system
CNAP
®
Monitor 500
Page 6-2
Version 5.2.5,
h
21-FHZU-10002
6.1
Visual alarm signals
Alarm signals are displayed visually in the
Alarm frame
and the
Parameter frame
,
directly in
the main screen. In the
Parameter frame
, only those parameters are visually marked which
exceed the defined alarm limits.
Illustration 27: Visual alarm signals
Alarm frame:
•
Background color:
YELLOW - Physiological alarms (medium priority)
CYAN - Technical malfunction alarms (low priority)
BLUE - Status messages
•
Alarm priority:
**
- medium priority
*
- low priority
•
Alarm system status:
- Alarm
- Alarms Paused
- Alarms Off
- Alarm Acknowledged timed
- Alarm Acknowledged infinite
•
Alarm message: a text with an alarm message describing the cause for the alarm signal appears
in the
Alarm frame
.
•
Flashing
Alarm frame
for physiological alarms
Distributed by BIOPAC Systems, Inc.
Manufactured by CNSystems AG