Management of patient data
CNAP
®
Monitor 500
Page 5-4
Version 5.2.5,
h
21-FHZU-10002
Units:
In the menu Setup | Monitor | Units it is possible to select the unit system for entering the patient
data – the options are Metric and US Customary.
Property
Metric
US Customary
weight
kg
lb
height
m
ft
and
in
5.3
Discharge
As a rule, patient data need to be deleted when a measurement in a new patient is performed:
The
Discharge
function
•
deletes all information in the
Patient Data
menu,
•
deletes all trends of data from the monitor,
•
deletes all entries of the
Alarm History
.
After having stopped the measuring process, patient data can be deleted in 2 ways:
a)
Patient Data
menu:
Open the
Patient Data
menu by using the click-wheel control to select the
Alarm
frame
on the main screen. Select
Discharge
and confirm your selection in the input di-
alog.
b)
Setup of a new patient:
Immediately after a new patient has been set up with the CNAP
®
double finger cuff and the
device’s self-test has been performed, the
Setup Patient
dialog appears on the main
screen. Select
New Patient – Pediatric Defaults
or
New Patient – Adult De-
faults
and confirm your choice in the input dialog. This will result in the deletion of any
previous patient data.
NOTE:
•
In order to avoid loss of patient data, all required data and entries must be printed
before discharging (=deleting patient data) a patient.
•
Discharge
can only be made after the measurement has stopped.
•
After
Discharge
,
even before leaving the
Patient Data
menu, new patient data
can be entered for the next patient. In such a case, this patient data will be used in
the following
Setup Patient
dialog by selecting
Use Current Patient Data
.
Distributed by BIOPAC Systems, Inc.
Manufactured by CNSystems AG