Cleaning and disinfection
CNAP
®
Monitor 500
Page 13-1
Version 5.2.5,
h
21-FHZU-10002
13
Cleaning and disinfection
Only use disinfectants and detergents recommended by CNSystems Medizintechnik AG to clean or
disinfect the device and its accessories. CNSystems’ warranty does not cover any damage caused
by the use of unsuitable cleaning agents or methods.
The warranty of CNSystems does not apply to the effectiveness of the mentioned cleaning agents
and methods for the purpose of infection prevention and control. When in doubt, the operator
should contact the hospital hygiene department. This particularly applies for the effectiveness of
disinfectants and detergents against hepatitis B and HI viruses. The operator is to follow the regu-
lations of the respective hospital and country.
13.1
General precautions
The CNAP
®
Monitor 500 including all its components and accessories are to be kept clean and free
of dust. After cleaning and disinfecting the devices, they must be thoroughly inspected before use.
If any components show signs of wear or damage, these components must not be used for patient
measurements! Before sending devices and components back to CNSystems Medizintechnik AG,
they are to be decontaminated.
CAUTION:
•
Always dilute detergents and disinfectants according to manufacturers' instruction, or
use in the smallest possible concentration.
•
No liquid must ingress the CNAP
®
Monitor 500.
•
Do not dip instruments, device parts or components in liquid.
•
Do not pour any liquid directly on the device.
•
Do not let residues of detergents air-dry on any parts of the device. Wipe them off
with a cloth moist with water, and then dry the instruments with a clean cloth.
•
Never use scouring agents or abrasive detergents (e.g. steel wool or silver polish).
•
Do not use bleaching agents!
•
Wipe off detergents with a moist cloth (water), then dry surfaces with a clean cloth.
STOP:
No liquid must be spilt on any part of the CNAP
®
Monitor 500. In case this should hap-
pen, carefully dry device/accessory. If in doubt whether liquid has entered the device, do
not start up the instrument. Contact technical staff or a service partner of CNSystems
Medizintechnik AG.
13.2
Cleaning
In order to clean any part of the device use a lint-free cloth, moist with warm water (max. 40° C),
and soap, diluted non-caustic detergents, tensides or detergents containing ammonia or alcohol.
Do not use strong solvents like dimethylketone or trichloroethylene. Do not dip the device, any part
of the device or any accessories (especially not the hoses) into liquid.
As the screen of the CNAP
®
Monitor 500 is easily scratched, be particularly careful when cleaning
it. No liquid must enter the CNAP
®
Monitor 500, so be sure to not spill any liquid directly on the
monitor. No liquid must enter the connectors of the CNAP
®
Monitor 500 or the CNAP
®
controller, so
take care not to wipe over, but rather around the connectors when cleaning them.
CAUTION:
Be particularly careful when cleaning or disinfecting the insides of the CNAP
®
double
finger cuffs. Wipe them carefully in order to avoid any damage.