Monitor configuration
CNAP
®
Monitor 500
Page 4-2
Version 5.2.5,
h
21-FHZU-10002
Parameter
Averaging
Averaging of displayed numeric parameters -
Default: 5 beats
Off, 5, 10, 15 beats
Patient Category
Presetting of patient category as a focus in the
Setup Patient
dialog -
Default: Adult
Adult, Pediatric
NOTE:
•
Measurement settings are saved automatically for any current or future measure-
ment.
•
Loss of power supply (interruption of mains operation, followed by depletion of
battery) entails the loss of measurement settings.
•
All settings can be changed to factory settings in the
Service
menu (see chapter
4.3 – Service settings.
The following table gives an overview of the three different NBP modes:
•
Trending Calibration: Immediately after the start of a measurement the CNAP
®
signal is ca-
librated via a transfer function. This trending calibration is intended to display values until
another calibration (NBP or external) is applied.
•
Height Offset: Starting from the trending calibration the user can enter a height offset be-
tween heart level and finger sensor – negative values mean that the finger sensor is below
heart level.
•
External Calibration: In this mode the CNAP
®
Monitor expects a manually entered blood
pressure value (e.g. measured with a NBP module from a patient monitor) for calibrating
the CNAP
®
signal.
•
NBP Calibration: The CNAP
®
Monitor calibrates automatically to a blood pressure value
measured with the internal NBP module.
The default setting for the NBP Measurement Type can be applied in the menu
Setup | Measure-
ment | NBP Measurement Type
. For the current measurement the measurement type can be
changed in the Start Measurement dialog and the menu
Parameters | NBP Measurement Type.
4.2.1
Trending calibration
The trending calibration is applied automatically and is intended for intermediate calibration until
the CNAP
®
Signal is calibrated to an internal NBP measurement or an external blood pressure val-
ue. In trending mode the CNAP
®
values are displayed with a gray background color and the mes-
sage “Waiting for BP Calibration” is displayed in the calibration info field.
NOTE:
•
One precondition for the trending calibration is that the patient is in supine position
and the hand with the finger sensor on is on heart level. Otherwise high deviations
of the measurement signal may be possible.
CAUTION:
•
It is highly recommended to calibrate the CNAP signal to a internal or external NBP
measurement before performing any blood pressure therapy on a patient.