![CNSystems CNAP Monitor 500 HD Скачать руководство пользователя страница 114](http://html1.mh-extra.com/html/cnsystems/cnap-monitor-500-hd/cnap-monitor-500-hd_operators-manual_2628121114.webp)
Appendix C
CNAP
®
Monitor 500
Page 17-3
Version 5.2.5,
h
21-FHZU-10002
17.1.2
CNAP
®
- continuous non-invasive arterial pressure
CNAP
®
- continuous non-invasive arterial pressure
Parameter classification
Sys, Dia, MAP [mmHg]
Pulse [bpm]
Measuring range
Sys:
40 - 250 mmHg (5.3 – 33.3 kPa)
Dia:
30 - 210 mmHg (4 - 28 kPa)
MAP:
35 - 230 mmHg (4 – 30.6 kPa)
Heart rate indication range
30-200 bpm
Accuracy
±5 mmHg (0.6 kPa)
Display resolution
1 mmHg (0.1 kPa)
Inflation pressure
Typ.: 120 mmHg (16 kPa)
Min.:
30 mmHg (4 kPa)
Max.: 300 ±10 mmHg (41.3 kPa ±1.3 kPa)
Excess pressure limit
300 ±10 mmHg (40 kPa ±1.3 kPa)
Response time: < 3 sec.
Deflation time: < 15 sec
Protection against electric
shock
Defibrillation protected applied part – Type BF
Defibrillation protection
According to IEC 60601-1
HF surgery
No restrictions
17.1.3
NBP - non-invasive blood pressure
NBP - non-invasive blood pressure
Parameter classification
Sys, Dia [mmHg]
Measuring method
Oscillometric: diastolic value for phase 5 Korotkoff
Measuring range
Sys:
ADULT
40 - 260 mmHg
PEDIATRIC
40 - 230 mmHg
Dia:
ADULT
20 - 200 mmHg
PEDIATRIC
20 - 160 mmHg
Heart rate indication range
30-220 bpm
Inflation pressure at start
ADULT:
160 mmHg
PEDIATRIC:
140 mmHg
Clinical accuracy
Meets EN 1060-4; ANSI/AAMI/ISO 81060-2
Accuracy of pressure
recording
± 3mmHg between 0 - 300 mmHg at operating temperatures of 0 –
50°C
Calibration interval for
pressure recording
12 months
Distributed by BIOPAC Systems, Inc.
Manufactured by CNSystems AG