Technical alarms and status messages
CNAP
®
Monitor 500
Page 14-8
Version 5.2.5,
h
21-FHZU-10002
Message
Priority
Possible cause
Measures
CNAP: Cuff Fault –
Unlicensed
Low *
-
CNAP
®
double finger
cuff is not licensed
for CNAP
®
Monitor
500
-
Check for permutation with
equipment from third
party
devices
CNAP: Finger Cuff
Expired
Low *
-
CNAP
®
double finger
cuff has reached end
of lifecycle, thus
providing low quality
of measurement – it
must be replaced
immediately
-
Replace CNAP
®
double finger
cuff immediately
-
Order new CNAP
®
double
finger cuff in corresponding
size
CNAP: Cuff Fault –
Leakage Left
Low *
-
Leakage in left
CNAP
®
finger cuff
-
Check connections of CNAP
®
hardware
-
Replace CNAP
®
double finger
cuff (check with other cuff
size)
-
In case of persistent error,
replace CNAP
®
double finger
cuff
CNAP: Cuff Fault –
Leakage Right
Low *
-
Leakage in right
CNAP
®
double finger
cuff
-
Check connections of CNAP
®
hardware
-
Replace CNAP
®
double finger
cuff (check with other cuff
size)
-
In case of persistent error,
replace CNAP
®
double finger
cuff
CNAP: Cuff Failure –
Inflation Timeout
Low *
-
Inflation of CNAP
®
double finger cuff
exceeded time limit
-
Disconnect CNAP
®
hardware
-
Reboot CNAP
®
monitor
-
In case of persistent error,
contact service
CNAP: Controller
Fault – Memory
Low *
-
Memory chip in
CNAP
®
controller de-
fective
-
Disconnect and reconnect
CNAP
®
controller
-
In case of persistent error,
contact service for faulty
CNAP
®
controller
CNAP: Controller
Fault – Unlicensed
Low *
-
CNAP
®
controller is
not licensed for
CNAP
®
Monitor 500
-
Check for permutation with
equipment from third
party
devices