Appendix C
CNAP
®
Monitor 500
Page 17-6
Version 5.2.5,
h
21-FHZU-10002
17.3
Electromagnetic compatibility
Medical electric devices have to comply with special safety regulations regarding EMC (electromag-
netic compatibility). Please keep in mind the respective precautions in this operator's manual be-
fore installing and operating the CNAP
®
Monitor 500.
Also, pay attention to the fact that portable and mobile HF-communication devices (e.g. mobile
phones) may interfere with medical electric devices.
The CNAP
®
Monitor 500 must not be placed directly beside or stockpiled with other devices. If
there is no other way but to operate the CNAP
®
Monitor 500 directly beside or stockpiled with other
devices, the CNAP
®
Monitor 500 must be closely observed to ensure its normal operability within
this arrangement of devices.
Only original CNSystems Medizintechnik AG accessories and power cords are to be used with this
device! Authorized accessories and replacement parts are listed in "Appendix B – Accessories" in
this operator's manual. Using third party manufacturer accessories may result in increased elec-
tromagnetic emission or in decreased functional immunity of the CNAP
®
Monitor 500.
As electric and magnetic fields may interfere with the functional reliability of the device, avoid us-
ing the CNAP
®
Monitor 500 close to devices emitting powerful electromagnetic fields, e.g. x-ray
equipment, diathermy applications or magnetic resonance tomographs.
Guidelines and manufacturer's declaration – electromagnetic emissions
The CNAP
®
Monitor 500 is intended for use in an electromagnetic environment as specified below. The cus-
tomer or operator of the CNAP
®
Monitor 500 is to ensure it is used in such an environment.
Emission test
Compliance
Electromagnetic environment -
guidelines
RF emissions
CISPR 11
Group 1
The CNAP
®
Monitor 500 uses RF ener-
gy only for its internal function. There-
fore, its RF emissions are very low and
are not likely to cause any interfe-
rence in nearby electronic equipment.
RF emissions
CISPR 11
Class B
The CNAP
®
Monitor 500 is suitable for
use in all establishments, including
domestic establishments and those
directly connected to the public low
voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
NA
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
NA
Distributed by BIOPAC Systems, Inc.
Manufactured by CNSystems AG