29
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
Disconnect the pendant from the unit before
attempting any removal or replacement procedures
to prevent electrical shock.
6 REMOVAL & REPLACEMENT
6.12 PENDANT BATTERY
A. Part Number
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89870
B. Tools and Equipment Required
#1 Phillips Screwdriver
Soldering Iron
Solder
NOTE:
Refer to
Section B
of the section referenced
below for additional tools needed.
C. Pendant Battery Removal & Replacement
•
•
•
1.
Refer to
Section 6.10, part C, Pendant Cable
Removal and Replacement
to open the
Pendant and remove the Cable.
2. Hold the sides of the Battery. While holding
touch one of the solder points securing the
Battery to the Control Board and lift to detach
that side of the Battery from the Control
Board.
Refer to Figure 6.25.
3. Repeat this process on the second solder
point.
4. To install the Battery, use a soldering iron
and solder to place a small bead of solder on
each of the solder points on the Control
Board.
5. Hold the Battery on the beads of solder
placed on the Control Board.
6. Use a soldering iron to heat the beads of
solder securing the Battery to the Control
Board.
FIGURE 6.25
BATTERY
SOLDERING POINTS
Содержание OPTIFLEX3 2090
Страница 1: ...Model 2090 Applies to Serial numbers 1000 and above SERVICEMANUAL ISO 13485 CERTIFIED ...
Страница 38: ...36 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS ...
Страница 39: ...37 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS CONTINUED ...
Страница 44: ...42 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 1 LANDSCAPED ...
Страница 45: ...43 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 2 LANDSCAPED ...
Страница 46: ...44 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 PENDANT LANDSCAPED ...