OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
2 SAFETY PRECAUTIONS
2.2 PRECAUTIONARY INSTRUCTIONS (CONTINUED)
DO NOT connect the unit to an electrical
supply without first verifying that the
power supply is the correct voltage.
Incorrect voltage may cause unit damage,
malfunction, electrical shock, fire, or
personal injury. Your unit was constructed
to operate only on the electrical voltage
specified on the Voltage Rating and Serial
Number Plate. Contact your Chattanooga
Group dealer if the unit is not properly
rated.
Power Supplies retain High Voltage!
Materials that have been in contact
with bodily fluids must be handled
and disposed of in accordance with
National, Local, and Facility disposal rules,
regulations, and procedures.
Exercise caution when using accessories
and auxiliary devices such as muscle
stimulators, cold packs and other
modalities. Route lead wires, hoses,
tubes, etc. away from the working
mechanism of the OptiFlex3 to help
prevent damage to the OptiFlex3 and
any other modality used with it.
Unconscious patients or patients under
heavy influence of medication must be
constantly attended and monitored while
the OptiFlex3 is in use.
The OptiFlex3 must be completely visible
at all times during use. Never cover the
unit with bedding or any other means of
concealment while in operation.
If the OptiFlex3 is used in conjunction
with the optional OptiFlex “T” Trolley,
make certain the OptiFlex3 unit is
resting on the mattress of the bed and
the OptiFlex “T” is suspended with no
weight on the casters to prevent possible
movement of the unit and possible injury
to patient.
This unit must not be operated with any
adapter attached to the three prong plug
that would disable the earth connection.
Disruption of the earth connection may
cause unit damage, malfunction, electrical
shock, fire, or personal injury.
•
•
•
•
•
•
•
•
Failure to re-install and properly tighten all screws
may result in Electrical Safety System degradation
which may cause unit damage, malfunction,
electrical shock or personal injury.
•
Содержание OPTIFLEX3 2090
Страница 1: ...Model 2090 Applies to Serial numbers 1000 and above SERVICEMANUAL ISO 13485 CERTIFIED ...
Страница 38: ...36 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS ...
Страница 39: ...37 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS CONTINUED ...
Страница 44: ...42 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 1 LANDSCAPED ...
Страница 45: ...43 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 2 LANDSCAPED ...
Страница 46: ...44 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 PENDANT LANDSCAPED ...