13
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
5.7 FLEXION ANGLE & CALIBRATION TEST
Specification
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90° ± 3°
A. Equipment Required
Inclinometer, Protractor, or Goniometer accurate
to
1°.
B. Test
1.
Place unit on level work surface.
2.
Plug Power Cord into grounded wall outlet
with
appropriate
voltage.
See
Specifications
on
page 8.
3.
Connect Pendant to unit and turn unit
power
ON.
4.
Press and hold the Extension button. While
holding the Extension button down, press the
Down Arrow button until 0 is displayed on
the
Pendant.
Refer to Figure 5.2.
5.
Press and hold the Time and Extension
buttons. While holding the Time and
Extension buttons, press the Up or Down
Arrow button until 5 Sec is displayed on
the
Pendant.
Refer to Figure 5.3.
6.
Press and hold the Time and Flexion buttons.
While holding the Time and Flexion buttons,
press the Up or Down Arrow button until
5 Sec is displayed on the Pendant.
Refer to
Figure
5.4.
FIGURE 5.2
FIGURE 5.4
FIGURE 5.3
5 TROUBLESHOOTING
DOWN ARROW
BUTTON
EXTENSION
BUTTON
EXTENSION
BUTTON
TIME
BUTTON
DOWN ARROW
BUTTON
UP ARROW
BUTTON
FLEXION
BUTTON
TIME
BUTTON
DOWN ARROW
BUTTON
UP ARROW
BUTTON
Содержание OPTIFLEX3 2090
Страница 1: ...Model 2090 Applies to Serial numbers 1000 and above SERVICEMANUAL ISO 13485 CERTIFIED ...
Страница 38: ...36 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS ...
Страница 39: ...37 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS CONTINUED ...
Страница 44: ...42 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 1 LANDSCAPED ...
Страница 45: ...43 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 2 LANDSCAPED ...
Страница 46: ...44 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 PENDANT LANDSCAPED ...