14
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
5 TROUBLESHOOTING
7.
Press and hold the Flexion button. While
holding the Flexion button down, press the
Up Arrow button until 90° is displayed on
the
Pendant.
Refer to Figure 5.5.
8.
Press and hold the Speed button. While
holding the Speed button down, press the
Up or Down Arrow button until 90° /Min
is displayed on the Pendant.
Refer to
Figure
5.6.
9.
While continuing to hold the Speed Button,
press the Emergency Start/Stop button on the
Pendant. When movement of the unit stops,
press the Emergency Start/Stop button. Verify
the display reads 90° /Min.
10.
Using the Inclinometer or Protractor,
measure the angle of the flexion of the unit.
Refer to Figure 5.7.
NOTE:
In
Figure 5.7
, a Digital Inclinometer is used
to measure each side and the two numbers
are added together.
If angle is out of specified range, perform
7.1 Angle Potentiometer Position
Calibration.
FIGURE 5.7
FIGURE 5.6
FIGURE 5.5
FLEXION
BUTTON
UP ARROW
BUTTON
SPEED
BUTTON
DOWN ARROW
BUTTON
UP ARROW
BUTTON
Содержание OPTIFLEX3 2090
Страница 1: ...Model 2090 Applies to Serial numbers 1000 and above SERVICEMANUAL ISO 13485 CERTIFIED ...
Страница 38: ...36 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS ...
Страница 39: ...37 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS CONTINUED ...
Страница 44: ...42 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 1 LANDSCAPED ...
Страница 45: ...43 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 2 LANDSCAPED ...
Страница 46: ...44 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 PENDANT LANDSCAPED ...