OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
5 TROUBLESHOOTING
5.8 TRAVEL SPEED TEST
Spec
. . . . . . . . . . . . . . . . . . . . . . 0° to 90° in 60 Sec ±10%
of seconds
A. Equipment Required
Calibrated stop watch
B. Test
1.
Set unit up as described in the “Flexion Angle
Test."
2.
Adjust the Femur Bar length to 34 and
tighten.
3.
Move to 0° by pressing the Emergency
Start/Stop button. When motion stops and
Display reads 0°, press the Emergency Start/
Stop button to stop the unit. Then,
simultaneously press the Emergency Start/
Stop button and the stop watch start button.
The unit should move from 0° to 90°
within the specification listed above.
4.
Repeat three times. Calculate and Record
the
average.
5.
If unit fails test, troubleshoot the following
components and replace if necessary:
a.
Check Pendant Cable connector to Motor
Control Board
b.
Motor Control Board
c.
Pendant
d.
Motor Assembly
e.
Drive Belt
f.
Pulleys
5.9 EMERGENCY START/STOP FUNCTION TEST
Spec
. . . . . . . . .Unit start or stop upon pushing button
A. Equipment Required
OptiFlex3 with Pendant
B. Test
1.
Place unit on a level work surface.
2.
Plug unit power cord into grounded wall
outlet with appropriate voltage. See
Specifications
on
page 8.
3.
Connect Pendant to unit and turn unit power
ON.
4.
Press the Emergency Start/Stop button to start
and stop the unit. Repeat several times.
5.
If unit fails test, troubleshoot the following
components and replace if necessary:
a.
Pendant Cable connector to Motor
Control Board.
b.
Motor Control Board.
c.
Pendant.
5.10 PENDANT DISCONNECT TEST
Spec
. . . . . . . . . . . . . . . . . .Unit stops immediately
upon Pendant disconnect
A. Equipment Required
OptiFlex3 with Pendant
B. Test
1.
Place unit on a level work surface.
2.
Plug unit power cord into grounded wall
outlet with appropriate voltage. See
Specifications
on
page 8.
3.
Connect Pendant to unit and turn unit
power
ON.
4.
Press the Emergency Start/Stop button to
start the unit.
5.
Disconnect the Pendant from the unit while
moving. Unit should stop.
6.
If unit fails test, troubleshoot the following
components and replace if necessary:
a.
Pendant Cable connector to Motor
Control
Board
b.
Motor Control Board
c.
Pendant
Содержание OPTIFLEX3 2090
Страница 1: ...Model 2090 Applies to Serial numbers 1000 and above SERVICEMANUAL ISO 13485 CERTIFIED ...
Страница 38: ...36 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS ...
Страница 39: ...37 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS CONTINUED ...
Страница 44: ...42 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 1 LANDSCAPED ...
Страница 45: ...43 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 2 LANDSCAPED ...
Страница 46: ...44 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 PENDANT LANDSCAPED ...