OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
A. Part Number
. . . . . . . . . . . . . . . . . . . . . . . . . . . .J6086
B. Tools and Equipment Required
#1 Phillips Screwdriver
C. Pendant Cable Removal & Replacement
1. Disconnect the Pendant Cable from the
Pendant Cable Receptacle by turning the
Pendant Cable connector counterclockwise.
Continue to unscrew to release.
2. Turn the Pendant face down. Using a #1
Phillips screwdriver, loosen the 7 screws
securing the back Cover of the Pendant.
Refer
to
Figure
6.21.
•
Disconnect the pendant from the unit before
attempting any removal or replacement procedures
to prevent electrical shock.
FIGURE 6.21
NOTE:
The part number for the screws securing
the back of the pendant is J1120. The O-ring is
part number J1121.
3. Fold the Rubber Gasket on the handle of the
Pendant and lift the back Cover off to remove.
SCREWS
FOLD BACK
THIS GASKET
4. Detach the 5 pin connector from the Control
Board of the Pendant.
Refer to Figure 6.22.
FIGURE 6.22
5 PIN
CONNECTOR
RUBBER
GASKET
5. Lift the Rubber Gasket and pull to release the
Gasket and the Cable from the base of the
Pendant.
6. Replace in reverse order.
6 REMOVAL & REPLACEMENT
6.10 PENDANT CABLE
Содержание OPTIFLEX3 2090
Страница 1: ...Model 2090 Applies to Serial numbers 1000 and above SERVICEMANUAL ISO 13485 CERTIFIED ...
Страница 38: ...36 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS ...
Страница 39: ...37 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS CONTINUED ...
Страница 44: ...42 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 1 LANDSCAPED ...
Страница 45: ...43 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 2 LANDSCAPED ...
Страница 46: ...44 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 PENDANT LANDSCAPED ...