4
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
2 SAFETY PRECAUTIONS
2.2 PRECAUTIONARY INSTRUCTIONS
Read, understand and practice the precautionary and
operating instructions found in this manual before
operating or using the unit. Know the limitations
and hazards associated with using the OptiFlex3
Continuous Passive Motion (CPM) Therapy Unit.
Observe any and all precautionary and operational
decals placed on the unit.
Only use OptiFlex3 on firm, flat, level surfaces.
Extreme caution should be taken when in use with or
around children.
Use OptiFlex3 only for its intended purpose as
described in this manual.
Turn power switch off before unplugging unit from its
power source.
Do not use the cord to unplug the power cord from
the unit. Grasp at the power cord base.
Transport and store the OptiFlex3 in temperatures
between 0° and 140 °F (-18° to 60 °C) to prevent
damage to the unit or its components.
Use extra care when touching metal of OptiFlex3 after
exposure to cold or heat to prevent static shock to
persons and or the unit.
Condensation could result and damage OptiFlex3
if unit is subjected to periods of low temperatures
followed by periods of high temperatures.
Use care when carrying, transporting or storing the
OptiFlex3 unit to prevent damage to the unit from
dropping or improper transport and storage methods.
Keep hair, loose clothing, fingers and all parts of
the body except the limb being treated, away from
moving parts of the OptiFlex3.
Unplug the Power Cord when not in use.
Do not use Power Cords that are damaged or frayed.
OptiFlex3 is made from high impact materials.
However, structural failure or hidden damage can be
caused by shock, impact, or dropping the unit. Use
care when transporting and storing unit to avoid
equipment damage.
To isolate the unit from the power source, disconnect
the power cord at the wall outlet.
Rapid increases in ROM can cause complications.
Tool, lubrication, and locking compound
requirements are critical to component removal and
replacement of the OptiFlex3.
All hardware, bolts, nuts, and screws used to assemble
the OptiFlex3 are SAE Standard. Due to the size of
these components, no metric equivalent is available.
Therefore, it will be necessary to obtain the proper
size tools for removal and replacement of certain
components.
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Federal law restricts this device to sale by, or on the
order of, a physician or licensed practitioner.
Make certain that the unit is electrically grounded
by connecting only to a grounded electrical service
receptacle conforming to the applicable national and
local electrical codes.
Keep hair, loose clothing, loose bedding, fingers and
toes away from the hinge components of the unit.
Do not use the OptiFlex3 outdoors or on wet surfaces.
Use only on firm, flat, stable level surfaces to ensure
stability of the unit while in operation.
Materials of the unit may become flammable or
combustible if exposed to a source of ignition.
Heat generated within the pendant may cause
ignition of the pendant if wrapped in bedding or other
materials.
Do not use OptiFlex3 while smoking or around open
flame.
OptiFlex3 has been designed for protection against
the exposure of urinary incontinence. Precautionary
measures should still be taken when any type of liquid
comes in contact with an electrical apparatus.
Always turn off and unplug unit from electrical source
before servicing or cleaning. Failure to do so could
result in electrical shock or personal injury.
Handle the unit only when unit is dry and hands are
dry to prevent electrical shock.
Do not use the OptiFlex3 as a toy.
A unit failing dielectric withstand and/or leakage tests
could indicate serious internal system problems. Do
not place unit back into service. Contact the factory for
repair. Do not attempt to repair the unit in the field.
Unplug the unit from the power source before
attempting any removal or replacement procedures to
prevent electrical shock.
•
•
•
•
•
•
•
•
•
•
•
•
•
The lubricants and locking compounds listed in
this manual are crucial in the assembly of certain
components to ensure patient safety and efficient
operation of the unit. Use only the recommended
products listed or an approved equivalent possessing
the same properties and qualities.
•
Содержание OPTIFLEX3 2090
Страница 1: ...Model 2090 Applies to Serial numbers 1000 and above SERVICEMANUAL ISO 13485 CERTIFIED ...
Страница 38: ...36 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS ...
Страница 39: ...37 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS CONTINUED ...
Страница 44: ...42 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 1 LANDSCAPED ...
Страница 45: ...43 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 2 LANDSCAPED ...
Страница 46: ...44 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 PENDANT LANDSCAPED ...