22
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
6 REMOVAL & REPLACEMENT
NOTE:
To order a replacement Potentiometer, order all
three parts listed in
section A.
A. Part Numbers
Potentiometer
. . . . . . . . . . . . . . . . . . . . . . . . . J6040
Split
Pin
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . J6023
Nylon
Covers
. . . . . . . . . . . . . . . . . . . . . . . . . . J6022
B. Tools and Equipment Required
3/32" Allen Wrench
#2 Phillips Screwdriver
Heyco Tool #29
Fine edged tool
Small Flathead Screwdriver
NOTE:
Refer to
Section B
of each sections
referenced below for additional tools needed.
C. Potentiometer Removal & Replacement
1. Refer to
Section 6.1, parts C and D
, Rear
Cover and Front Cover Removal and
Replacement to remove the Covers of the
unit.
2. Use a fine edged tool to lift the Decal
on the Potentiometer and remove.
Refer to
Figure
6.9.
NOTE:
Save the Decal, J6107, for installation of the
replacement
Potentiometer.
•
•
•
•
•
Unplug the unit from the power source before
attempting any removal or replacement procedures
to prevent electrical shock.
POTENTIOMETER
DECAL
COVER
SCREWS
3. Use a #2 Phillips screwdriver to remove the 3
screws securing the Potentiometer Cover and
remove the Cover.
NOTE:
When installing the Potentiometer Cover, part
number J6105, hand tighten the three
screws until snug. Overtightening the screws
will crack the Cover.
4. Using a 3/32" Allen wrench remove the two
cleat
screws.
Refer to Figure 6.10
5. Using a Heyco Tool, squeeze the sides of the
Heyco and pull out to remove.
Refer to
Figure
6.10.
6. Using a small Flathead screwdriver positioned
behind the Potentiometer, apply just enough
pressure to free the Potentiometer.
Refer to
Figure
6.10.
NOTE:
On the back of the Potentiometer, make
certain that the Nylon Covers, part number
J6022, on each side of the split pin are
installed.
Refer to Figure 6.11
SMALL
SCREWDRIVER
CLEAT
HEYCO
FIGURE 6.9
FIGURE 6.10
FIGURE 6.11
SPLIT PIN WITH
NYLON COVERS
Failure to re-install and properly tighten all
screws may result in Electrical Safety System
degradation which may cause unit damage,
malfunction, electrical shock or personal injury.
6.7 POTENTIOMETER (KNEE-POT)
Содержание OPTIFLEX3 2090
Страница 1: ...Model 2090 Applies to Serial numbers 1000 and above SERVICEMANUAL ISO 13485 CERTIFIED ...
Страница 38: ...36 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS ...
Страница 39: ...37 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS CONTINUED ...
Страница 44: ...42 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 1 LANDSCAPED ...
Страница 45: ...43 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 2 LANDSCAPED ...
Страница 46: ...44 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 PENDANT LANDSCAPED ...