16
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
6 REMOVAL & REPLACEMENT
6.1 UNIT COVERS
A. Part Number
Push Rivet (Male & Female Components). . J6004
Rear Cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . J6089
Front Access Cover. . . . . . . . . . . . . . . . . . . . . . . J6103
B. Tools and Equipment Required
Miniature Diagonal Cutters
C. Rear Cover Removal and Replacement
NOTE:
Miniature Diagonal Cutter are used only to
remove the Rivets. They are not intended to
cut the Rivets. Only apply enough pressure to
grasp the Rivet. They will be used to
re-install
the
Cover.
NOTE:
If Carriage is close to full extension, there may
not be sufficient clearance to remove the
Cover. For inoperable units,
refer to Section
6.2 Manual Movement of the Carriage
for
carriage
adjustment.
1. Using a pair of miniature diagonal cutters
remove the 11 Rivets securing the Rear or
Motor Cover. Grasp the Rivets from behind and
gently pry the Rivets up to release. This will
release both the male and female parts of the
Rivet.
Refer to Figure 6.1.
•
2. Lift up and rotate the Rear Acess Cover toward
the Carrying Handle of the unit. Slide the
cover over the Carrying Handle to completely
remove from the unit.
3. To install the Rear AcessCover, reverse the
above
steps.
D. Front Access Cover Removal and
Replacement
NOTE:
Miniature Diagonal Cutter are used only to
remove the Rivets. They are not intended to
cut the Rivets. Only apply enough pressure to
grasp the Rivet. They will be used to
re-install
the
Cover.
NOTE:
If Carriage is close to full extension, there may
not be sufficient clearance to remove the
Cover. For inoperable units,
refer to Section
6.2 Manual Movement of the Carriage
for
carriage
adjustment.
1. Using a pair of miniature diagonal cutters
remove the 8 Rivets securing the Rear or
Motor Cover. Grasp the Rivets from behind
and gently pry the Rivets up to release. This
will release both the male and female parts of
the
Rivet.
2. Lift to remove.
3. Reverse above steps to install.
Unplug the unit from the power source before
attempting any removal or replacement procedures
to prevent electrical shock.
FIGURE 6.1
Содержание OPTIFLEX3 2090
Страница 1: ...Model 2090 Applies to Serial numbers 1000 and above SERVICEMANUAL ISO 13485 CERTIFIED ...
Страница 38: ...36 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS ...
Страница 39: ...37 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS CONTINUED ...
Страница 44: ...42 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 1 LANDSCAPED ...
Страница 45: ...43 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 2 LANDSCAPED ...
Страница 46: ...44 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 PENDANT LANDSCAPED ...