12
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
5 TROUBLESHOOTING
5.3 VISUAL INSPECTION
A. General
Visually inspect the OptiFlex3 for and visible
damage to the unit or pendant that may interfere
with operation of the unit. Make sure that all
screws are secure. Verify that the appropriate hook
or loop fastener is secured to the carriage.
Check for proper labeling. Make sure that the
Femur adjustment decals are in good condition.
Verify that the serial decal for the unit is present.
Verify that the LCD lens is in good condition and
clean if necessary.
Verify that the domed heads of the femur hinge
rivets are in good condition. Make sure that
the hinges are secure, by attempting to pull them
apart.
5.4 FUNCTIONAL INSPECTION
Check the calibration of the unit by running it
successively through its full ROM with femur length
adjusted first to maximum and then to its
minimum
length.
NOTE:
Verify that the lead screw is unobstructed and no
abnormal noises are present during operation.
Disconnect the pendant while the unit is running
to verify that the unit will stop.
Perform the Speed test found on
page 15.
Clears all settings by activating the Reset.
5.5 ELECTRICAL SAFETY
The
OptiFlex3 has been tested to UL 60601-1,
Standard for Safety for Medical Equipment.
NOTE:
The device complies with current leakage, ground
continuity, and dielectric withstand (Hi-Pot) limits
as prescribed by IEC/EN/UL 60601-1 and C22.2 No.
601.1-M90 w/A2 Medical Electrical, Part 1: General
Requirements
for
Safety.
Facility, local and national limits and test methods
may
vary.
A. Power Requirements
Model: 2090 ..................................................... Input: 100-240 VAC
50/60 HZ, 75 VA
5.6 ELECTRICAL SAFETY TESTS
Conduct all necessary Electrical Safety tests as
required per your facility, local or national
regulatory agency. In the USA follow NFPA 99
(National Fire Protection Association) "Health care
Facility"
standards.
NOTE:
Use the rivet shown in
Figure 5.1
as a connection
point for Grounding Tests.
A unit failing dielectric withstand and/or leakage
tests could indicate serious internal system
problems. Do not place unit back into service.
Contact the factory for repair. Do not attempt to
repair the unit in the field.
This unit must not be operated with any
adapter attached to the three prong plug
that would disable the earth connection.
Disruption of the earth connection may cause
unit damage, malfunction, electrical shock, fire,
or personal injury.
Rivet
FIGURE 5.1
Содержание OPTIFLEX3 2090
Страница 1: ...Model 2090 Applies to Serial numbers 1000 and above SERVICEMANUAL ISO 13485 CERTIFIED ...
Страница 38: ...36 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS ...
Страница 39: ...37 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS CONTINUED ...
Страница 44: ...42 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 1 LANDSCAPED ...
Страница 45: ...43 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 2 LANDSCAPED ...
Страница 46: ...44 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 PENDANT LANDSCAPED ...