34
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
7 CALIBRATION
7.2 FORCE REVERSAL CALIBRATION
A. Position Carriage
Move the carriage to a position of -5° to -10°.
B. Set Unit to Force Reversal Calibration
Mode
While holding down the “SPEED” button,
press the following buttons in the sequence
listed:
1.
Progressive ROM
2.
Fast Back
3.
Oscillation Zone
4.
Mode
5.
Release the Speed button.
“STAND BACK! Unit Self Calibrating" should be
displayed at the bottom of the Pendant
display and a Timer should begin counting
down from approximately 30 sec.
Refer to
Figure
7.10.
FIGURE 7.10
6.
After all movement of the unit has stopped,
the Display will prompt the technician to turn
the unit off and back on.
Refer to Figure
7.11.
This completes Force Reversal
Calibration.
FIGURE 7.11