24
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
A. Part Number
. . . .
. . . . . . . . . . . . . . . . . . . . . . . . J2004
B. Tools and Equipment Required
5/64" Allen wrench
#2 Phillips screwdriver
Gear Puller
Snap Ring Pliers
5/16" Socket Driver
Large Flathead Screwdriver
Needle Nose Pliers
Molygraph Grease
NOTE:
Refer to
Section B
of each sections referenced
below for additional tools needed.
C. Acme Rod Removal & Replacement
1. Refer to
Section 6.1, parts C and D
, Rear
Cover and Front Cover Removal and
Replacement to remove the Covers of the
unit.
2. Refer to
Section 6.4, part C
, Drive Belt
Removal and Replacement, to remove the
Drive
Belt.
3. Using a 5/64" Allen wrench, remove the set
screw from the Pulley attached to the Ball
Screw.
NOTE:
Retain all rivets, nuts, washers, bearing and
screws for installation of the Ball Screw.
4. Use a gear puller to remove the Pulley from
the Ball Screw.
NOTE:
Using a gear puller should decrease the
possibility of damage to the Pulley.
5. Remove the 2 washers and the bearing
from the end of the Ball Screw.
Refer to
Figure
6.12.
•
•
•
•
•
•
•
•
7. Using a 5/16" Socket Driver remove the 4
nuts and washers securing the Base
Plate at the end of the Frame Base.
Refer to
Figure
6.14.
6. Using a #2 Phillips screwdriver, remove
the screw on each side of the Frame Base to
release the Base.
Refer to Figure 6.13.
SCREW SECURING BASE
TO THE FRAME BASE
FIGURE 6.13
FIGURE 6.14
8. Using a large Flathead screwdriver, gently
pry the plate off of the Ball Screw.
Refer to
Figure
6.14.
NUTS SECURING
BASE PLATE
PLACE FLATHEAD
SCREWDRIVER BEHIND
THIS PLATE
Unplug the unit from the power source before
attempting any removal or replacement procedures
to prevent electrical shock.
NOTE:
Observe the locations of the washers and
bearing for installation.
Inset
A on page 38
details proper installation.
WASHERS
BEARING
FIGURE 6.12
6 REMOVAL & REPLACEMENT
Failure to re-install and properly tighten all
screws may result in Electrical Safety System
degradation which may cause unit damage,
malfunction, electrical shock or personal injury.
6.8 ACME ROD (BALL SCREW)
Содержание OPTIFLEX3 2090
Страница 1: ...Model 2090 Applies to Serial numbers 1000 and above SERVICEMANUAL ISO 13485 CERTIFIED ...
Страница 38: ...36 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS ...
Страница 39: ...37 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit MAINTENANCE 8 4 MAINTENANCE RECORDS CONTINUED ...
Страница 44: ...42 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 1 LANDSCAPED ...
Страница 45: ...43 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 CONTROL BOARD 2 LANDSCAPED ...
Страница 46: ...44 OptiFlex 3 Continuous Passive Motion CPM Therapy Unit 10 SCHEMATICS 10 2 PENDANT LANDSCAPED ...