31
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
Disconnect the pendant from the unit before
attempting any removal or replacement procedures
to prevent electrical shock.
6 REMOVAL & REPLACEMENT
6.14 PENDANT DISPLAY
A. Part Number
. . . . . . . . . . . . . . . . . . . . . . . . . . .J1123
B. Tools and Equipment Required
#1 Phillips Screwdriver
NOTE:
Refer to
Section B
of the section referenced
below for additional tools needed.
C. Pendant Receptacle Cable Removal &
Replacement
1.
Refer to
Section 6.11, part C, Control Board
Removal and Replacement
to open the
Pendant, remove the Cable, and remove the
Control
Board.
2. Using a #1 Phillips screwdriver, remove the
4 screws securing the Pendant Display to the
Upper case of the Pendant.
Refer to
Figure
6.27.
•
FIGURE 6.27
SCREWS SECURING
DISPLAY
3. Lift the display to remove.
4. Reverse for installation.