![Caliper Life Sciences Twister II Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/caliper-life-sciences/twister-ii/twister-ii_user-manual_3442863037.webp)
Operation
37
P/N 113564 Rev. 2
Twister II Robot User’s Manual
Caliper Life Sciences, Inc.
Adding Materials to the Twister II Racks in iLink PRO
(Continued)
Figure 15. Twister II Output Rack
If the method discards the material to a waste location or puts it
into a different storage instrument
instead of placing it into an
empty rack, you do not need to define an output rack.
7
If a method uses
more than one material in the same Twister
II rack
, for example, plates with lids or 4 empty daughter plates,
drag the additional plates into the racks one at a time, following
steps 1 through 6 for each material.
If a method uses
more than one material in the same rack
,
the materials must physically be stacked in the order in which
the method uses the materials. The integrator must specify the
order of multiple materials in the same rack for operators. Only
the exact same plate type and part number can be in a rack,
with the exception of lids.
Example
: If you are placing 96 well plates into a rack, you may
only place that type of plate, and its lids, into that rack.
Specifically,
two plates with the same specifications but
from different manufacturers may not be used in the same
rack.
Daughter A and Daughter B plates, if they are in the same
rack, must be the same plate type, from the same manufacturer,
with the same part number.
See
“Example: Adding Multiple Materials to Twister II Racks” on
for a step-by-step procedure to add multiple plates to
multiple Twister II racks in the iLink PRO software.