Appendix A: Software Reference 151
P/N 113564 Rev. 2
Twister II Robot User’s Manual
Caliper Life Sciences, Inc.
Instrument Teach Wizard - Place Microplate Window
The Place Microplate window describes where to place the plate for
teaching and prepares to release the motors for teaching.
Figure 128. Instrument Teach Wizard, Place Microplate Window
Clicking
Next
causes the robot arm to drop. Be sure to hold the
robot arm before clicking Next to prevent the arm from dropping.
Instrument Teach Wizard - Teach Location 1 Window
The Teach Location 1 window shows how to position the robot and
plate for teaching the first instrument location.
Figure 129. Instrument Teach Wizard, Teach Location 1 Window
Clicking
Next
creates position 1 on the Position tab.
Grip buttons
- Open and close the gripper if you neeed to grip or
release a plate.
Servo Control buttons
- Enables or disables the robot axes.