Twister II - Maintenance 107
P/N 113564 Rev. 2
Twister II Robot User’s Manual
Caliper Life Sciences, Inc.
Changing the Twister II Fuse(s) (Continued)
c
Remove the 3Amp fuse.
d
Turn the fuse block over and replace with the two fuse
locations facing up, as shown in
e
Tighten the screw to secure the fuse block to the cover.
f
Place the two 1.6Amp fuses for 240V configuration into the
fuse block as shown in
6
Replace the fuses with the same part number fuses.
7
Confirm that the fuses are seated securely in the fuse block.
8
Snap the fuse block/cover assembly back into the housing. See