700 NDIR Analyzer
6-20-2017
89
Raw Values Display
→
→
→
or
→
The Raw Values Display screen is a diagnostic tool for viewing detector Raw Volts
and Calculated Concentrations.
This screen is accessed by pressing
from the
Diagnostics menu.
Raw Voltage:
This is a 0.512 VDC to 4.512 VDC detector output that will be digitized
by the microprocessor to generate the calibration curve from which the Raw
Concentration and Measured Concentration are derived. The 0.512 volts is equal to 0
ppm and 4.512 volts is equal to the highest factory-set range limit of that component.
Raw Concentration:
This value (in engineering units) is calculated from the Raw Volts
before linearization and offset and span corrections are applied.
Measured Concentration:
This value (in engineering units) is calculated from the Raw
Concentration after linearization, offset and span corrections have been applied.