03H Read Floating Point
Modbus Commands Use TCP/IP Port 502. (Do not change TCP/IP port from 7700.)
Register Numbers and Descriptions
Register Number
Contents IEEE
40001
UNDILUTED Channel 1 concentration = diluted conc.* dil. ratio / 10000
40003
DILUTED Channel 1 concentration
40005
Channel 1 concentration before linearization & zero / span corrections
40007
Channel 1 raw detector volts
40009
UNDILUTED Channe1 2 concentration = diluted conc.* dil. ratio / 10000
40011
DILUTED Channel 2 concentration
40013
Channel 2 concentration before linearization & zero / span corrections
40015
Channel 2 raw detector volts
40017
UNDILUTED Channel 3 concentration = diluted conc.* dil. Ratio / 10000
40019
DILUTED Channel 3 concentration
40021
Channel 3 concentration before linearization & zero / span corrections
40023
Channel 3 raw detector volts
40025
Channel 1 current range full scale concentration
40027
Channel 2 current range full scale concentration
40029
Channel 3 current range full scale concentration
40031
Flow 1
40033
Flow 2
40035
Flow 3
40037
External analog 1
40039
External analog 2
40041
Cell pressure
40043
Analyzer temperature
40045
Channel 1 detector temperature
40047
Channel 2 detector temperature
40049
Channel 3 detector temperature
40051
Channel 1 EPC voltage
Содержание 700 NDIR
Страница 1: ...700 NDIR Analyzer 6 20 2017 1 ...
Страница 32: ...700 NDIR Analyzer 6 20 2017 32 ...
Страница 99: ...700 NDIR Analyzer 6 20 2017 99 Comp Factors or ...
Страница 100: ...700 NDIR Analyzer 6 20 2017 100 T P Compensation H2O Compensation ...
Страница 101: ...700 NDIR Analyzer 6 20 2017 101 Gas Compensation ...