700 NDIR Analyzer
6-20-2017
133
Linear Coefficients
→
→
→
The Linear Coefficients function allows the operator to optimize linearity by inputting up
to five coefficients for each range to generate up to a fourth-order curve.
From the Service
menu press
to access the Linear Coefficients screen.
NOTES:
Changing linear coefficients can compromise the analyzer’s accuracy and ability to
function properly.
To reset Factory Coefficients see Reset Factory Settings.
Select the corresponding function key for the range you wish to edit. In the example,
was
pressed to allow the Range 1 Linear Coefficient to be changed.
Содержание 700 NDIR
Страница 1: ...700 NDIR Analyzer 6 20 2017 1 ...
Страница 32: ...700 NDIR Analyzer 6 20 2017 32 ...
Страница 99: ...700 NDIR Analyzer 6 20 2017 99 Comp Factors or ...
Страница 100: ...700 NDIR Analyzer 6 20 2017 100 T P Compensation H2O Compensation ...
Страница 101: ...700 NDIR Analyzer 6 20 2017 101 Gas Compensation ...