700 NDIR Analyzer
6-20-2017
44
Introduce zero gas into the rear of the analyzer. Press
to go to the
Diagnostic
Values
screen
to view the current diagnostic values. Check the temperatures and flows to
be sure they are within their limits. If all diagnostic values are within their limits, press
the
button to return to the Manual Zero Calibration screen.
When the concentration value has stabilized, press
to set the zero calibration. The
zero portion of the calibration should now be complete.
If the calibration was successful, the screen will say ***Saved Current*** above the
concentration value.
Example of a successful calibration.