700 NDIR Analyzer
6-20-2017
202
16H Write Floating Point
Modbus Commands Use TCP/IP Port 502. (Do not change TCP/IP port from 7700.)
Register Numbers and Descriptions
Register Number
Contents IEEE
40201
Channel 1 Range 1 span gas concentration
40203
Channel 1 Range 2 span gas concentration
40205
Channel 1 Range 3 span gas concentration
40207
Channel 1 Range 4 span gas concentration
40209
Channel 2 Range 1 span gas concentration
40211
Channel 2 Range 2 span gas concentration
40213
Channel 2 Range 3 span gas concentration
40215
Channel 2 Range4 span gas concentration
40217
Channel 3 Range 1 span gas concentration
40219
Channel 3 Range 2 span gas concentration
40221
Channe1 3 Range 3 span gas concentration
40223
Channel 3 Range 4 span gas concentration
40225
Dilution ratio
40227
Channel 1 flow alarm minimum
40229
Channel 1 flow alarm maximum
40231
Channel 2 flow alarm minimum
40233
Channel 2 flow alarm maximum
40235
Channel 3 flow alarm minimum
40237
Channel 3 flow alarm maximum
40239
External analog 1 minimum
40241
External analog 1 maximum
40243
External analog 2 minimum
40245
External analog 2 maximum
40247
Cell pressure minimum
40249
Cell pressure maximum