700 NDIR Analyzer
6-20-2017
109
Once the output type has been selected, use the Left/Right arrows
to move the
highlight into the % FS column of the output to be checked. Press
until the % FS
value reads 0.000. To adjust the
zero (Offset), use the Left/Right arrows
to
highlight the Offset column and press
to open the field. Make a small adjustment to
the Offset value and then press
to close the field. Repeat this procedure as
necessary.
To adjust the output to Full Scale, use the Left/Right arrows
to move the
highlight into the % FS column and press
until the % FS column value reads 100.0.
Then use the Left/Right arrows
to highlight the Gain column. Press
to
open the field. Make a small adjustment and press
again to close the field and check
the output. Repeat this procedure as necessary for other outputs.
When you are finished making all your changes press
to save them. You will return
to the Programmable Analogs menu.
NOTE:
Failure to save your adjustments will result in the numbers reverting back to the
defaults after power is turned off and back on.
The following table includes typical Programmable Analog Output values:
OUTPUT
OFFSET
GAIN
0-20 mA
0.000
0.828
4-20 mA
1.658
0.662
0-1 V
0.828
0.828
0-5 V
0.828
0.828
0-10 V
0.828
0.828
Содержание 700 NDIR
Страница 1: ...700 NDIR Analyzer 6 20 2017 1 ...
Страница 32: ...700 NDIR Analyzer 6 20 2017 32 ...
Страница 99: ...700 NDIR Analyzer 6 20 2017 99 Comp Factors or ...
Страница 100: ...700 NDIR Analyzer 6 20 2017 100 T P Compensation H2O Compensation ...
Страница 101: ...700 NDIR Analyzer 6 20 2017 101 Gas Compensation ...