700 NDIR Analyzer
6-20-2017
108
Output Adjustments
→
→
→
→
→
The Output Adjustments screen allows the operator to set the output to either mA
or voltage and calibrate the outputs to exactly match the results obtained on a PLC
or other remote data-recording device.
Using the Output Adjustments screen, the
operator can force the analog outputs to 0%, 50% or 100% of Full Scale and back to
Measure. For information on analog output connections see
Analog and Digital Interface
.
From the Programmable Analogs menu press
to access the Output Adjustments
screen.
Output Type
can be set for mA or 1, 5 or 10 VDC.
Output AO-1
refers to analog output 1.
% FS
is used to toggle between Measurement, 0%, 50% and 100% Full Scale.
Offset
(zero) is used to adjust the output at 0%.
Gain
(span) is used to adjust the output at 100%.
To select the
Output Type
press
to open the highlighted field. Use the Up/Down
arrows
to make your selection. Press
again to close the field.