Analyzer Specifications
Specifications are subject to change without notice.
IR Analysis Method:
Non-dispersive infrared (NDIR)
NDIR Components:
CO, CO
2
, CH
4
, SO
2
, NO
Detector Type:
Microflow
NDIR Ranges*:
From 0-50 ppm up to 0-100% *SO
2
, 0-1% up to 0-20%, and CH
4
0-
1,000 ppm up to 0-100%
Range Ratio:
10:1
Response Time (IR):
90% of Full Scale in < 1 second, depending on cell length, flow
rate and time constant
IR Sample Cell:
Stainless steel w/ replaceable gold cell liner
Resolution:
Typically 0.1% of Full Scale
Repeatability:
Better than 1.0% of Full Scale
Linearity:
Better than 1.0% of Full Scale of factory-calibrated ranges
Noise:
Less than 1% of Full Scale of factory-calibrated ranges
Zero and Span Drift:
Less than 1% of Full Scale per 24 hours
Sample Flow Rate:
0.5 to 2.0 LPM
Oxygen Analysis Method:
Paramagnetic or Electrochemical Fuel Cell
O
2
Ranges:
0-1% (paramagnetic only) up to 0-100% O
2
Full Scale; four definable ranges
O
2
Response Time:
T90 -
2
seconds paramagnetic; 16 seconds fuel cell
Outputs Available:
Voltage, current, TCP/IP, RS-232, Modbus
Display:
3” x 5” backlit LCD
Sample Temperature:
Up to 50°C, non-condensing
Ambient Temperature:
5 to 40°C
Ambient Humidity:
Less than 90% RH (non-condensing)
Fittings:
¼-inch tube
Power Requirements:
115/230 (
±
10%) VAC; 50/60Hz; 300 Watts maximum
Dimensions:
5¼”H x 19”W x 23”D
Weight:
30-45 lbs., depending on configuration
Содержание 700 NDIR
Страница 1: ...700 NDIR Analyzer 6 20 2017 1 ...
Страница 32: ...700 NDIR Analyzer 6 20 2017 32 ...
Страница 99: ...700 NDIR Analyzer 6 20 2017 99 Comp Factors or ...
Страница 100: ...700 NDIR Analyzer 6 20 2017 100 T P Compensation H2O Compensation ...
Страница 101: ...700 NDIR Analyzer 6 20 2017 101 Gas Compensation ...