700 NDIR Analyzer
6-20-2017
193
Modbus Map
(NDIR) – V1 2/10/15
01H Single-Read Coil
Modbus Commands Use TCP/IP Port 502. (Do not change TCP/IP port from 7700.)
Coil Numbers and Descriptions
Coil Number
Read Data
1
Check Flow 1
2
Check Flow 2
3
Check Flow 3
4
Check External Analog 1
5
Check External Analog 2
6
Check pressure
7
Check temperature
8
Channel 1 is not calibrated
9
Channel 2 is not calibrated
10
Channel 3 is not calibrated
11
Channel 1: Low concentration warning
12
Channel 2: Low concentration warning
13
Channel 3: Low concentration warning
14
Channel 1: High concentration warning
15
Channel 2: High concentration warning
16
Channel 3: High concentration warning
17
Check Channel 1: Temperature
18
Check Channel 2: Temperature
19
Check Channel 3: Temperature
20
Check Channel 1: EPC voltage
21
Check Channel 2: EPC voltage
22
Check Channel 3: EPC voltage
23
Check Channel 1: Range overflow
24
Check Channel 2: Range overflow
25
Check Channel 3: Range overflow