700 NDIR Analyzer
6-20-2017
42
Manual Calibration
→
→
→
or
→
Whether you are calibrating a single range or multiple ranges, each range requires
its own complete zero and span calibration.
If you are calibrating multiple ranges
during one session, the zero calibrations can all be performed before any of the span
calibrations, as long as they are within the same relatively short time period. If multiple
ranges are used, the calibrations are typically done in ascending order of range. Anytime
a zero calibration is performed, a span calibration or check should be done afterward. The
Manual Calibration menu is accessed by pressing
from the Calibration menu.
Press
to access the Manual Zero Calibration screen.
Press
to access the Manual Span Calibration screen.