Diagnostic Values
→
→
→
or
→
Temperatures Screen
The Diagnostic Values screens allow the operator to check analyzer temperatures,
pressures, voltages and flows.
These important screens are accessed by pressing
from the Diagnostics menu.
The first screen that appears is the Temperatures screen. The Temperatures screen
displays the current temperature and low and high alarm limits for key analyzer
components.
As indicated at the bottom of the screen, use the left and right arrow keys
to
scroll to different screens. The current screen will be highlighted (Temps in the example).
The Temperatures, Pressures and EPC Voltage Percent screens include the current device
Values and the Low and High Alarm Limits. For example, if the CO detector temperature
drops lower than 45°C or exceeds 55
°
C an alarm will be triggered and displayed at the
bottom of the
Measure Screen
.