700 NDIR Analyzer
6-20-2017
169
EDAL: Set Diagnostic Alarm Limits
Continued
Command
Response
Description
_EDAL_K0_ai.min_
a1.max_..._a16max
_EDAL_s
Set all alarm limits.
11: Channel 1 detector temperature
12: Channel 2 detector temperature
13: Channel 3 detector temperature
14: Channel 1 sample EPC voltage %
15: Channel 2 sample EPC voltage %
16: Channel 3 sample EPC voltage %
_EDAL_K0_x_
x.min_xmax
_EDAL_s
Set alarm limit of x.
EFDA: Set Auto Calibration and Purge Times
Command
Response
Description
_EFDA_Km_SATK_
z_y_x
_EFDA_s
Set auto calibration times for channel m in
seconds.
z: Purge time
y: Verify time
x: Purge after
_ EFDA_K0_
SSPL_z
_EFDA_s
Set analyzer purge time to z seconds.
EGRD: Set the User Level’s Range Polynomial Coefficients
Command
Response
Description
_EGRD_Km_Mn_
A0_a1_a2_a3_a4
_EGRD_s
Set the user level’s polynomial coefficients for
channel m range n.
EGRW: Set Maximum Absolute and Relative Deviation Limits
Command
Response
Description
_EGRW_Km_Mn_z_y
_EGRW_s
Sets absolute and relative deviations for
channel m range n in %.
z: Absolute
y: Relative